Restriction
If you are using the concentration measurement application, you must set Volume Flow Type to Liquid. Gas
standard volume measurement is incompatible with the concentration measurement application.
Procedure
Set Volume Flow Type to Gas Standard Volume.
5.4.2
Configure Standard Gas Density
Display
Not available
ProLink II
ProLink→Configuration→Flow→Std Gas Density
Field Communicator
Configure→Manual Setup→Measurements→Gas Standard Volume→Gas Density
Standard Gas Density is used to convert the measured flow data to reference (standard) values.
Prerequisites
Ensure that Density Measurement Unit is set to the units you will use for Standard Gas Density.
Procedure
Enter the appropriate Standard Gas Density value for the gas you are measuring.
Tip
ProLink II provides a Gas Wizard that you can use to calculate the standard density of your gas, if you
do not know it.
5.4.3
Configure Gas Standard Volume Flow Measurement Unit
Display
OFF-LINE MAINT→OFF-LINE CONFG→UNITS→VOL
ProLink II
ProLink→Configuration→Flow→Std Gas Vol Flow Units
Field Communicator
Configure→Manual Setup→Measurements→Gas Standard Volume→Gas Vol Flow Unit
Gas Standard Volume Flow Measurement Unit specifies the unit will be used for the gas standard volume
flow rate. The unit used for the gas standard volume total and the gas standard volume inventory is
derived from this unit.
Prerequisites
Before you configure Gas Standard Volume Flow Measurement Unit, be sure that Volume Flow Type is set to Gas
Standard Volume.
Procedure
Set Gas Standard Volume Flow Measurement Unit to the desired unit.
The default setting for Gas Standard Volume Flow Measurement Unit is SCFM (standard cubic feet per minute).
Configuration and Use Manual
Configure process measurement
75