20 Setup
3.
Push it into place until it clicks.
4.
Reattach the cover. Use the alignment tab and slide the cover to the left to click it back into
place.

Remove the temperature probe and well

Follow these steps to disconnect the probe cable and remove the probe well.
1.
Remove the cover of the temperature module by pressing the tab and sliding the cover to the
right. The cover is located at the bottom-right of the monitor, below the probe well.
2.
Depress the spring tab on the temperature probe cable connector and withdraw it from the
probe port.
3.
Reattach the cover. Use the alignment tab and slide the cover to the left to click it back into
place.
Grasp the probe well and pull it up to remove it from the monitor.

Connect the NIBP hose

1.
Place your thumb and forefinger on the hose connector and squeeze the side tabs.
2.
Align the hose connector with the hose connector port on the side of the monitor.
Welch Allyn Vital Signs Monitor 6000 Series™