Remove The Battery - Welch Allyn Connex 6000 Series Service Manual

Vital signs monitor
Hide thumbs Also See for Connex 6000 Series:
Table of Contents

Advertisement

Service manual
Note
4. If the Power down options dialog appears, select the desired option and then touch
Power down.

Remove the battery

Before removing the battery, power down the monitor as described in "Power down the
monitor."
1. Turn the monitor upside down to access the battery cover.
2. Locate the battery cover, indicated by
3. Insert a coin into the slot and push to open. Choose a coin that fits comfortably into
the slot.
4. Pull the battery out by pulling the battery label, which is visible when you open the
battery cover.
If the monitor is in the Continuous Monitoring profile, a
Power down options dialog appears with the following
options:
Save patient data on the device with a temporary Patient ID
Remove patient data from the device
Disassembly and repair 57
.