82 Patient monitoring
Set Response Mode
To set the Response Mode from the Home tab, the monitor must be in the Monitor profile.
Touch
The SpO2 frame will display
SatSeconds™ alarm management
The SatSeconds feature is an SpO2 alarm management system available only with monitors that
are equipped with Nellcor OxiMax Technology.
The SatSeconds feature is the product of the time and magnitude that a patient falls outside of the
SpO2 alarm limits. For example, three points below the alarm limit for 10 seconds equals 30
SatSeconds. An alarm is triggered only when a desaturation event reaches the SatSeconds limit.
The SatSeconds feature is clinician controlled and can be set to 0, 10, 25, 50, or 100 SatSeconds. If
a desaturation event resolves on its own within the preset time, the clock will automatically reset
and the monitor will not alarm.
Note
Set SatSeconds limits
1.
Touch the Alarm limit control of the SpO2 frame
2.
Touch the Alarms tab.
3.
Touch
4.
Touch Home to save your settings and return to the Home tab.
Perfusion index
Perfusion Index (PI) is an SpO2 feature available only with Masimo-equipped monitors.
in the SpO2 frame.
in the lower left corner when Fast mode is selected.
The SatSeconds feature has a built-in safety protocol that sounds an alarm
whenever three SpO2 violations of any amount or duration occur within a 1-minute
period.
to select a SatSeconds setting.
Welch Allyn Vital Signs Monitor 6000 Series™
.