94 Specifications
Protection classifications, all monitor configurations
Pixel arrangement
Color depth
Speaker volume
Output sound pressure
Alarm and pulse tones
Pulse frequency (f
Number of harmonic components in
the range 300 Hz to 4000 Hz
Effective pulse duration (t
Rise time (t
a
Fall time
Note
a
Prevents overlap of pulses.
Battery specifications
Rating
Composition
Charge time to 80 percent capacity
Charge time to 100 percent capacity
Patient exams per charge
Age to 70 percent capacity
1
A patient exam includes NIBP, Temperature, and SpO2 measurements at the rate of one patient every 10
minutes with a 2-minute display time out setting and a new battery.
2
After this many full charge and discharge cycles, the battery has aged to where its total capacity has been
reduced to 70 percent of its rating.
Nurse Call connection specifications
Nurse Call
)
0
)
d
)
r
(t
)
f
The relative sound pressure level of the harmonic components should be within 15 dB above
or below the amplitude at the pulse frequency.
1
2
Welch Allyn Vital Signs Monitor 6000 Series™
RGB (red, green, blue)
16 bits per pixel
67 dB at 1.0 meter
per IEC 60601-1-8
150 – 1000 Hz
minimum of 4
high priority: 75 –200 ms
medium and low priority: 125 – 250 ms
10 – 20% of t
d
t
< t
– t
f
s
r
6 cell
11.1 V 3.80Ah (42Wh)
Lithium-ion
2hr 7m
3hr
26
300
25 V AC or 60 V DC maximum at 1A maximum
9 cell
10.8 V 6.75Ah (73Wh)
Lithium-ion
2hr 25m
4hr
47
300