108 Guidance and manufacturer's declaration
Recommended separation distances between portable and mobile RF
communications equipment and the monitor
Rated max.
output power of
transmitter (W)
0.01
0.1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation distanced
in meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is
the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
Note 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects, and people.
150 kHz to 80
80 MHz to 800
MHz
MHz
d= (1.17)
d= (1.17)
0.11667
0.11667
0.36894
0.36894
1.1667
1.1667
3.6894
3.6894
11.667
11.667
Welch Allyn Vital Signs Monitor 6000 Series™
800 MHz to 2.5 GHz
d= (2.23)
0.23333
0.73785
2.3333
7.3785
23.3333