Fetch Subsystem - Chroma 62000H Series Operating & Programming Manual

Programmable dc power supply
Hide thumbs Also See for 62000H Series:
Table of Contents

Advertisement

Return Parameter: <NRf+> [Unit = Volt]
Query Example:
Return Example:
Note: The output must be OFF for setting.

5.6.2.4 FETCH Subsystem

1.
FETCh:VOLTage?
Description:
Query Syntax:
Return Parameter: <NRf+> [Unit Volt]
Query Example:
Return Example:
2.
FETCh:CURRent?
Description:
current.
Query Syntax:
Return Parameter: <NRf+> [Unit Amp]
Query Example:
Return Example:
3.
FETCh:POWer?
Description:
Query Syntax:
Return Parameter: <NRf+> [Unit Amp]
Query Example:
Return Example:
4.
FETCh:STATus?
Description:
Query Syntax:
Return Parameter: <Arg1><,><Arg2><,><Arg3>
SOUR:DCON:FALL?
SOUR:DCON:RISE?
7.950000e+01
It measures the output of Power Supply and returns real time
voltage.
FETCh:VOLTage?
FETC:VOLT?
9.983100e+00
It measures the output of Power Supply and returns real time
FETCh:CURRent?
FETC:CURR?
2.000000e-04
It measures the output of Power Supply and returns real time power.
FETCh:POWer?
FETC:POW?
5.000000e+03
It returns the status code of Power Supply's state.
FETCh:STATus?
<Arg1>: return warning message 0~65535, 0: no warning, use
binary for the rest and identify the cause of error.
BIT 0:
OVP
BIT 1:
OCP
BIT 2:
OPP
BIT 3:
Remote Inhibit
BIT 4:
OTP
BIT 5:
FAN_LOCK
BIT 6:
SENSE FAULT
BIT 7:
SERIES FAULT
BIT 8:
Reserved
BIT 9:
AC FAULT
BIT 10: FOLD Back CV to CC
BIT 11: FOLD Back CC to CV
BIT 12: Reserved
BIT 13: Reserved
Remote Operation
It returns the setting.
5-19

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents