Current Measurement Calibration - Chroma 62000H Series Operating & Programming Manual

Programmable dc power supply
Hide thumbs Also See for 62000H Series:
Table of Contents

Advertisement

9.
Now the voltage calibration is done. To save the calibration data, press "
prompt a confirmation page as Figure 3-95 shows. Press "
set SAVE=YES and press "
"
EXIT
" to return to the Calibration screen.
10. Press "
" to return to the MAIN PAGE.
EXIT
1.
2.

3.3.7.2 Current Measurement Calibration

3.3.7.2.1 Hardware Requirements
Table 3-7 lists the hardware requirements for current measurement calibration.
DVM
CURRENT SHUNT
ELECTRICAL LOAD
LOAD
BREAKER
Figure 3-94
" to save it. If there is no need to save it, press
ENTER
Figure 3-95
The calibration point may be different for other models (non
62150H-600), please operate it following the instructions displayed.
It is necessary to remove the output load when performing voltage
calibration. The LCD panel will show the text as Figure 3-93 and
once no load is confirmed for the output, press "
calibration.
Table 3-7
Device
Suggest Model or Capacity
HP 34401A or equivalent DVM
Prodigit 7550 or equivalent
CHROMA 63204 or equivalent
Capable current>=100A
Manual Operation
" will
SAVE
1
" or "Rotary" (
" to start
ENTER
) to
3-53