As to the voltage setting, the inputted analog voltage 0~5V maps to the actual output voltage
0~1200V; and for the current setting, the inputted analog voltage 0~5V maps to the actual
output current 0~25A as Figure 3-51(a) shows. Set the APG option to APG VSET/APG
ISET = Vref(0-10V) means the inputted analog voltage 0~10V maps to the actual output
0~1200V for APG voltage also maps to the actual output 0~25A for APG current as Figure
3-51(b) shows. For the above voltage/current setting method, the analog voltage (0~5V or
0~10V) has to be inputted to the devices connected in series respectively in order to have
the effect of serial operation in APG mode.
1200V/25A
3.3.3.6.2 Parallel Setting
To connect 5 sets of 62150H-600 DC Power Supplies in parallel for operation and set the
APG option to APG VSET/APG ISET = Vref(0-5V), the MAIN PAGE of MASTER will show as
Figure 3-52.
As to the voltage setting, the inputted analog voltage 0~5V maps to the actual output voltage
0~600V; and for the current setting, the inputted analog voltage 0~5V maps to the actual
output current 0~125A as Figure 3-53(a) shows. Set the APG option to APG VSET/APG
ISET = Vref(0-10V) means the inputted analog voltage 0~10V maps to the actual output
0~600V for APG voltage also maps to the actual output 0~125A for APG current as Figure
3-53(b) shows. For the above voltage/current setting method, the analog voltage (0~5V or
0~10V) has to be inputted to the devices connected in series respectively in order to have
the effect of serial operation in APG mode.
APG INPUT 5V/5V
(a)
Figure 3-51
Figure 3-52
1200V/25A
APG INPUT 10V/10V
(b)
Manual Operation