Chroma 62000H Series Operating & Programming Manual page 96

Programmable dc power supply
Hide thumbs Also See for 62000H Series:
Table of Contents

Advertisement

Programmable DC Power Supply 62000H Series Operating & Programming Manual
The table above lists the BREAKER capacity for 62150H-600 only. The
CAUTION
applicable BREAKER for other models, please see the specifications of
OUTPUT CURRENT in Table 1-1 to Table 1-3.
3.3.7.2.2 SETUP
Figure 3-96 is the diagram for connecting current calibration devices.
62000H
DC Power Supply
When conducting the current calibration, each calibration point has to key
in at least 5 Arabic numerals to ensure the Power Supply accuracy after
calibration.
3.3.7.2.3 Calibration Procedure (Example: Model 62150H-600)
1.
Enter into the page of Figure 3-90.
2.
In CALIBRATION page, press "
3.
Press "
" to confirm and entering into current calibration options as Figure 3-97
ENTER
shows.
4.
Open the Breaker to ensure the DC Power Supply has no load and press "
confirm.
5.
It is important to reconnect the DC Power Supply to current shunt whose rating is
closest to but still cover 2A. For Prodigit 7550, use 2A shunt directly.
3-54
Current Shunt
+
sv
+
v
v
Breaker
sv
Figure 3-96
2
" or turn "Rotary" (
Figure 3-97
+
) knob to set CHOICE=2.
+
DVM
+
Electronic Load
" to
ENTER

Hide quick links: