confirm. Use 0.5A and 1.5A for calibration, the system will calculate the calibration factor
for the low current range.
8.
Now start the high current range calibration. Set the current shunt whose rating is
closest to but still cover 10A. For Prodigit 7550, use 10A shunt
9.
Press "
" will perform 2.5A calibration. The system will set the output current to
2.500A automatically and the cursor will stop at position [3] as Figure 3-108 shows.
Input the current read by Current Shunt (DVM) to position [3] and press "
confirm.
10. The DC Power Supply will be set to off again. Set the current shunt whose rating is
closest to but still cover 10A. For Prodigit 7550, use 10A shunt directly.
11. Press "
" will perform 6.5A calibration. The system will set the output current to
6.500A automatically and the cursor will stop at position [4] as Figure 3-109 shows.
Input the current read by Current Shunt (DVM) to position [4] and press "
confirm.
Figure 3-107
Figure 3-108
Figure 3-109
Manual Operation
directly.
" to
ENTER
ENTER
" to
3-59