Cleaning and Disinfection Instructions ........26 Battery Instructions ..............28 Care and Preventive Maintenance ..........29 Troubleshooting ..............33 Technical Specifications ............34 Dimensions ................36 Labels ..................38 Approvals and List of Standards ..........40 Electromagnetic Compatibility ..........40 Parts and Accessories ............42 ArjoHuntleigh Addresses ..........Last page...
Thank You for Purchasing ArjoHuntleigh Equipment. If a DVD accompanies this IFU, please note that the Your Carendo shower and toilet chair is part of a series DVD does not replace the information in this IFU. The of quality products designed especially for hospitals, IFU contains additional important safety instructions.
The equipment is intended for indoor use. maintenance in this IFU. The Carendo must be used by appropriately trained Depending on level of wear some equipment parts such caregivers with the adequate knowledge of the care...
(ISO 10993). the body and use designated grab supports during any movement. NOTE Do not lift the Carendo shower and toilet chair using the armrests; use the chassis. WARNING Only one function movement (e.g reclined function) can be done at a time.
Actions Before the First Use (4 steps) (9 steps) 1 Check that all parts of the Carendo shower and 1 Visually check the package for damage. If the toilet chair are in place. Compare to section Parts product looks damaged make a complaint to the Designation on page 8.
Product Description / Functions Reclined Function Fig. 1 A function which makes hygienic care of the patient easier. From a sitting position, a reclined position can be achieved by means of turning gears which makes the backrest move in backwards and upright directions. Corresponding to this movement the leg rest is moving up and down.
(See Fig. 2) Castors Fig. 3 The Carendo shower and toilet chair is equipped with lockable castors. The castors will brake by pressing the castor lock tab down with your foot (A). (See Fig. 3) The castors are released by pressing up the centre tab (B).
Hand Control Fig. 1 Connect the connector of the hand control on the right side of the backrest. (See Fig. 1) The hand control has a hook on the back, making it easy to attach Fig. 2 it to various parts of the chair. (See Fig. 2) Hand control Fig.
Battery Charger Fig. 1 The battery holder should be attached to a suitable wall and the charger connected to a properly grounded wall socket. (See Fig. 1) NOTE The battery charger is designed for use in a dry location and should not be installed in bathrooms, etc. See NEA 8000 Battery Charger IFU.
Emergency Lowering Fig. 1 In case of electrical malfunction, the chair can be lowered by pushing the green button. (See Fig. 1) NOTE This function should only be used in emergency. Emergency Stop Fig. 2 In any case of emergency, all functions can be stopped by pushing the red button (A).
Bedpan and Bedpan Holder Fig. 1 WARNING To avoid entrapment or pinching of genitals, make sure there is enough clearance during movement over the tub edge, toilet, bedpan or other furniture. A bedpan holder can be mounted under the seat. (See Fig. 1-2) See separate instruction Bedpan Holder.
1 Apply the castor brakes on the wheelchair. 2 Lower the Carendo shower and toilet chair so that its seat is at the same level as the wheelchair seat, or as close as possible.
Follow the separate operating instructions regarding the passive lift. 1 Apply the brakes on the Carendo shower and toilet chair. 2 Place the passive lift with the patient as close as possible to the Carendo shower and toilet chair. Position the patient in the sling directly over the seat of the Carendo shower and toilet chair.
Page 18
Carendo shower and toilet chair. (See Fig. 2) 3 Apply the brakes on the active lift. 4 Lower the patient till he/she reaches the seat of the Carendo Fig. 3 shower and toilet chair. Unhook sling and remove, if present.
Transfer to the Shower (18 steps) Fig. 1 1 Place the Carendo shower and toilet chair with the patient in the shower room. 2 Apply the castor brakes. 3 Raise the chair in a sitting position to a comfortable height for the caregiver.
Page 21
16 Put on the incontinence pads, if used, and pull up the trousers completely. (See Fig. 7) 17 Adjust the Carendo shower and toilet chair back to a sitting position and lower the chair. Secure the patient with the safety belt, if needed.
Transfer to the Toilet (13 steps) The Carendo shower and toilet chair can be used regardless of whether the toilet is attached to the wall or to the floor. See the measurements below on matching toilet chairs: min 5.5 in min 140 16.9 in...
Page 23
11 Put on the incontinence pads, if used, and pull up the Fig. 5 trousers completely. (See Fig. 5) 12 Adjust the Carendo shower and toilet chair back to a sitting position and lower the chair. WARNING To avoid entrapment, make sure to keep the patient’s hair, arms and feet close to the body and...
Home Care Use Using the Carendo For Distributors and Renters Before using the Carendo shower and toilet chair in When transporting the Carendo shower and toilet home environment make sure that: chair to the customer with a van: (See Fig. 1-3) •...
Arjo Clean If you have any questions regarding disinfecting the • Arjo General Purpose Disinfectant Carendo shower and toilet chair, or wish to order • Arjo Disinfectant Cleanser IV disinfectant, contact ArjoHuntleigh Customer Service. See section Parts and Accessories on page 42.
Page 27
2 Clean 10 Allow a disinfection time according to the • the Carendo shower and toilet chair (start from instructions on the disinfectant bottle label. the top and move downwards) 11 Spray or shower water (approx. 25 °C (77 °F)) and •...
To avoid bodily injury, do NOT crush, puncture, open, dismantle or otherwise • If the Carendo shower and toilet chair is used mechanically interfere with the battery. frequently, it is recommended to use two batteries, one in operation and one as a stand-by in the •...
Care and Preventive Maintenance Caregiver Obligations The Carendo shower and toilet chair is subject to wear and tear, and the following actions must be performed when specified to make sure that the product remains within its original manufacturing specification. WARNING To avoid malfunction resulting in injury, make sure to conduct regular inspections and follow the recommended maintenance schedule.
Page 30
IFU. Every day Fig. 1 • Disinfect. The Carendo shower and toilet chair has to be disinfected immediately after every use. ArjoHuntleigh disinfectants and shampoos should be used in recommended concentrations. See section Cleaning and Disinfection Instructions on page 26.
Page 31
Every sixth months Fig. 7 Lubricate the locking pins. Use grease Wekem WS267 (can be purchased from ArjoHuntleigh). (See Fig. 7) Yearly The Carendo shower and toilet chair has to be serviced according to the table QUALIFIED PERSONNEL Action/Check on page 32.
Page 32
WARNING To avoid injury and/or unsafe product, the maintenance activities must be carried out at the correct frequency by qualified personnel using correct tools, parts and knowledge of procedure. Qualified personnel must have documented training in maintenance of this device. QUALIFIED PERSONNEL Action/Check Every YEAR Lubricate bearings...
Troubleshooting NOTE If the product does not work as intended, immediately contact your local ArjoHuntleigh representative for support. PROBLEM ACTION It is only possible to lower the lift. The lifting device of this equipment is equipped with a a safety nut.
Technical Specifications Carendo, Lift and Shower Chair Model BIB2000-01 Safe Working Load (SWL) (Maximum total load) 136 kg (300 lbs) Weight 64 kg (141 lbs) Maximum total weight (lift + patient) 200 kg (441 lbs) Chassis, maximum width 640 mm (25 1/4”) Width between legs 508 mm (20”)
Page 35
This requirement applies in the European Union. Do not dispose these products as unsorted municipal waste. You can return your device and accessories to ArjoHuntleigh or you can contact your local authorities for advice on disposal.
Dimensions Measurements in mm and in inches (") Sitting, lowest position Sitting, highest position Reclined, lowest position Reclined, highest position Care Raise, highest position Care Raise, lowest position...
Page 37
Measurements in mm and in inches (") Ø1530 (60 1/4”) Ø960 ( 5/8”) Turning radius, sitting Turning radius, reclined...
Labels Label explanation NEA0100-083 Product no. of the battery Attention Important user information for 24 Vdc 2.5 Ah Label correct use. Mandatory to read the Instructions Approval States approvals for the current for Use Label product. Battery Label States safety and environmental information for the battery.
Page 39
Attention Symbol Battery Label Approval Label Identification Label Emergency Lowering Label Data Label...
EMC information below. The Carendo shower and toilet chair is intended for use in the electromagnetic environment specified below. The customer or the user of the Carendo shower and toilet chair should assure that it is used in such an environment. WARNING Stacking or placing other electrical equipment next to this device is not recommended, it can interfere with the equipment’s operation and safety.
Page 41
RF communications equipment (transmitters) and the Carendo shower and toilet chair by maintaining a distance of not less than 1 meter to any part of the Carendo shower and toilet chair, including cables, if the transmitters output power rating exceeds 1 W.
Parts and Accessories For disinfectant contact Back Pillow Neck Pillow the ArjoHuntleigh local representative Bed pan & Holder Safety belt Battery and Charger...
ArjoHuntleigh Addresses AUSTRALIA FRANCE POLSKA ArjoHuntleigh Pty Ltd ArjoHuntleigh SAS ArjoHuntleigh Polska Sp. z o.o. 78, Forsyth Street 2 Avenue Alcide de Gasperi ul. Ks Piotra Wawrzyniaka 2 O’Connor CS 70133 PL-62-052 KOMORNIKI (Pozna ) AU-6163 Western Australia FR-59436 RONCQ CEDEX...
Page 44
Getinge Group is a leading global provider of products and systems that contribute to quality enhancement and cost ef ciency within healthcare and life sciences. We operate under the three brands of ArjoHuntleigh, Getinge and Maquet. Getinge provides solutions for infection control within healthcare and contamination prevention within life sciences. Maquet specializes in solutions, therapies and products for surgical interventions, interventional cardiology and intensive care.