Setting the location and fuel

Before carrying out measurements, the location and fuel must be
correctly selected, see Address/Location, page 32
and see Fuels, page 44.
6.2.2.

Flue gas

To achieve usable measurement results, the measurement
period of a flue gas measurement should be approx. 3 min
and the measuring instrument should display stable
measuring values.
Call up function:
[
]
Measurement options
1.
2. Select the fuel → [OK].
Carrying out the measurement:
1. Start measurement:
If a separate measurement of CO undiluted has not yet
been carried out, this value is calculated using the
measured values of the flue gas probe and is updated
continuously.
If CO undiluted and/or a draught measurement has already
been carried out separately, the value obtained is applied.
-
The measured values are displayed.
2. End measurement:
Options
>
[Options]
→ Clipboard: data is saved to the clipboard.
>
[Options]
is deleted.
>
[Options]
→ Save: the measured values are saved in a record.
>
[Options]
displayed as a flue gas matrix, see below.
>
[Options]
measured values per display page.
>
[Options]
→ Recalibrate: the gas sensors are set to zero.
>
[Options]
during a measurement): the measured value display menu is
opened.
www.
[OK]
[
].
[
].
Delete
clipboard: any data saved to the clipboard
Fluegas
matrix: the measured values are
Number of
lines: change the number of
Measurement
view: (This function is not available
Testo-Direct
6 Using the product
Flue Gas
→ [OK].
47