7 Maintaining the product
4. Remove the adapter (4) and hose (5), push the thermocouple
out of its holder and disconnect the cable (6) from the
thermocouple.
5. Connect cables to the new thermocouple (white -, green +) (7)
and push the thermocouple back into the holder.
6. Connect the adapter and hose to the thermocouple (8). Adjust
cables and hoses (9), refit halfshell elements.
7. Replace the probe shaft and lock the sealing cap by tightening
it anti-clockwise (as far as it will go, observe markings).
8. Replace halfshell handles and secure with screws.