6 Using the product
2. Enable CO measurement parameter: [OK].
3.
[Edit]
→ Enter the test gas concentration (nominal value).
4. Attach the connecting line of the test gas bottle to the
calibration adapter.
5. Apply test gas to the sensor.
6. Start recalibration: [Start].
7. Apply the target value once the actual value is stable
(adjustment): [OK].
-or-
Cancel (no adjustment): [esc].
8. Save changes: [Finished].
6.1.4.

Fuels

The fuel can be selected. The fuel-specific coefficients and limits
can be set.
In order to maintain the measuring accuracy of the
instrument, the correct fuel must be selected or configured.
Correct representation of measuring results is only
assured if the threshold values for the ideal range of the
corresponding measurement task have been set correctly.
The pre-set threshold values are typical values for the
selected system type and the chosen type of fuel.
Call up function:
>
[
]
Fuels
Activating fuels:
> Select the fuel → [OK].
-
The fuel is activated and the main menu is opened.
Setting coefficients:
1. Select the fuel → [Coeff.].
2. Select the coefficients: [Edit].
Possibly:
> Enter the password:
[OK].
3. Set values → [OK].
4. Save changes: [Finished].
Setting limits:
1. Select limit → [Edit].
2. Set values → [OK].
3. Save changes: [Finished].