Construction Requirements - Sirona SINIUS Installation Requirements

Dental system
Hide thumbs Also See for SINIUS:
Table of Contents

Advertisement

Sirona Dental Systems GmbH
Installation Requirements SINIUS / SINIUS CS / SINIUS TS
Connections: Media (see On-site installation [ → 9])
● Water supply
Pipe 10x1 mm, corner valve outlet 3/8"
● Compressed air supply line
Pipe 10x1 mm, corner valve outlet 3/8"
● Suction line
DN 40 HT-PP ISO 8283-3, inner diameter 36.5 mm
● Water drainage
DN 40 HT-PP ISO 8283-3, inner diameter 36.5 mm
● Installation pipe (power supplies)
min. DN 40 HT-PP ISO 8283-3, inner diameter 40 mm
● Installation pipe (IT)
DIN 40 HT-PP ISO 8283-3 (or corresponding flat conduit)
Connections: Electrical (see On-site installation [ → 9])
● Power cable: 3 x 1.5 mm
● Type B automatic circuit breaker
230 V AC, 16 A slow-blow or
● Type B automatic circuit breaker
100-115 V AC, 20 A slow-blow
● Suction machine control cable and call cable: 3 x 1.5 mm
● Wireless systems in 2.4 GHz frequency range available?
(e.g. room monitoring systems, video transmitters, etc.)
● If yes, this may damage the wireless foot control.
Please consult our Product Service team.
Uneven floors: Mounting plates (see Mounting plates [ → 38])
● No plate required (unevenness max. 2 mm, shimming plates can be used)
● Adapter plate, REF 63 07 511
● Demo plate (demo operation at a trade show), REF 63 08 717
● Load capacity of the floor is indicated (pressure and tensile loads).
● The treatment center can be safely anchored in the load-bearing structure (concrete/wood;
NOT screed).
63 22 668 D3561
D3561.021.01.06.02 12.2014
6.2

Construction requirements

Electrical connections for Teneo, Sinius
²
Uneven floors for Teneo, Sinius
²
☐ Yes
6 Checklist
6.2 Construction requirements
☐ No
53