Sirona Dental Systems GmbH
Installation Requirements SINIUS / SINIUS CS / SINIUS TS
63 22 668 D3561
D3561.021.01.06.02 12.2014
4.6

Technical data

Model designation:
Model designation, power ratings
Power connection:
Nominel current:
Type of earthing system:
Overvoltage category:
Average power consumption
(for dimensioning an air
conditioning system):
Power consumption in
Standby mode:
Main building fuse:
Protection class:
Unit classifications, without HF surgery
Device class in accordance
with Directive 93/42/EEC:
Degree of protection against
electric shock:
Degree of protection against
ingress of water:
Mode of operation
Mode of operation
4 Dimensions, technical data
4.6 Technical data
SINIUS / SINIUS CS / SINIUS TS
100 - 230 V AC ± 10%
50/60 Hz
2.2 A at 230 V
4.35 A at 115 V
5.0 A at 100 V
also max. 6 A for external devices
TN-C-S system or TN-S system
(according to IEC 60364-1)
2 acc. to IEC 60664-1
0.25 kW
3 W (without internal mini PC)
Automatic circuit breaker type B
100 – 115 V AC: 20 A slow burn
230 V AC: 16 A slow burn
Class I equipment
Class IIa equipment
Type B applied parts
Except SiroCam AF intraoral camera.
This is
Applied part type BF
Ordinary equipment (without protection
against ingress of water)
The foot control has an IP X1 degree of
protection against liquids (drip-proof).
Continuous operation with intermittent
loading corresponding to the dental
mode of working.
Drive motors for chair operation:
intermittent use, max. 2 minutes on and
18 minutes off
Permanently connected unit. Operation
is not permitted in mobile vehicles.
41