Sirona SINIUS Installation Requirements page 16

Dental system
Hide thumbs Also See for SINIUS:
Table of Contents

Advertisement

3 On-site installation
3.4 Underfloor installation of supply lines
16
Installation Requirements SINIUS / SINIUS CS / SINIUS TS
1
Water supply
Pipe 10x1 mm, corner valve outlet 3/8"
2
Compressed air supply
Pipe 10x1 mm, corner valve outlet 3/8"
3
Suction line
DN40 HT-PP ISO 8283-3
(polypropylene, inner diameter 36.5 mm)
4
Water drainage
DN40 HT-PP ISO 8283-3
(polypropylene, inner diameter 36.5 mm)
5
Installation pipe
DN40 HT-PP ISO 8283-3
(polypropylene, inner diameter 40 mm)
6
Control cable to relays for the
suction machine (
%
call cable (
),
Special function (#)
2
3 x 1.5 mm
(quality as in the power cable)
7
Power supply
2
3x1.5 mm
Circuit breaker: for 230 VAC: 16 A slow-blow
for 100 - 115 VAC: 20 A slow-blow
Recommended: Type B automatic circuit breaker
8
not connected
9
Installation pipe, inner diameter 50 mm
(or corresponding flat conduit) for the PC connection
Sirona Dental Systems GmbH
),
63 22 668 D3561
D3561.021.01.06.02 12.2014