wtw PhotoLab Spektral Operating Instructions Manual page 123

Table of Contents

Advertisement

Sample printout:
Graphical display of the characteristics
graphic: Pressing
displays the characteristics of
the user-defined method graphically.
27:05:99
306
TEST1
c [mg/l]
1
100
2
250
3
400
4
550
5
700
...
zero point:
slope (A/c):
corr. coefficient:
13 User-defined methods
11:33:09
COD
A (585 nm)
0.137
0.214
0.290
0.369
0.445
0.090
0.000496
0.997
The graphic shows the course of the absorbance in
the defined measuring range. In addition, the slope S,
zero point A0 and correlation coefficient r2 are dis-
played.
The current cursor position is displayed on a black
background.
and
move the cursor on the straight
line.
print: Print using
LQ300 printer with printer cable required
(accessories).
table: Return to the table using
In the table, the
record that was previously marked in the
graphic.
.
sign appears on the data
93