10.2.2 Measuring the kinetics

The sample warms up during kinetic mea-
surements with long measuring durations.
After the measurement is finished:
10 Special functions
→ Call up the measure kinetics menu item in the
cinetics menu.
→ Insert the cell or use
ment.
The message measuring ... appears.
The measuring process appears as a graphic with the
specified parameters.
The kinetic measurements are performed continually
at the specified intervals.
The following information is given in the upper display
area (from left to right):
the time until the next single measurement (the
countdown runs)
the present duration of the measurement (the time
runs)
the number of the last single measurement
the wavelength
the last measured value
The cursor is always positioned on the last measured
value.
stop: Terminate the measuring process and display the
results in a graphic (see below) using
return: Return to the kinetics menu using
After the measurement is finished, the photometer au-
tomatically changes to the graphical display of the ki-
netics.
to start the measure-
.
.
75