The photometer operates at an environ-
mental temperature of + 5 °C to + 35 °C.
During transport from cold to warm sur-
roundings, condensation can form result-
ing in the malfunction of the meter. Before
putting the photometer into service, wait
until it has adapted to the new environ-
mental conditions (see also chapter
19 T
).
ECHNICAL DATA
3.1
Preparing the photometer
Caution
Place the photometer on a hard, flat surface
and protect it against intensive light and
heat.
The Memory Card contains all data relevant
for the operation of the photometer.
3 Commissioning
The photometer is a sensitive measuring
instrument. Take this into consideration
when choosing the surroundings. Espe-
cially avoid vapors and gases that might
attack mechanical, optical and electroni-
cal components of the photometer.
Do not do anything to the photometer that
you would not do to your hightec camera.
→ Insert the Memory Card
ing at the photometer up to the stop. The label
with the type designation and software version
must point upwards.
into the lateral open-