pHotoFlex
STD
Socket assignment
Output format, ASCII
ba75958d05
10/2017
3
1 RTS
2 RxD
2
4
6
3 TxD
5
1
4 DTR
5 SG
6 CTS
RS 232
REC

4.7.2 Configuring the RS232 interface

For an error-free data transmission, the RS232 interface should be set
to the same transmission speed (Baud rate) for the pHotoFlex
and PC/printer.
You can set the following values for the baud rate on the
®
pHotoFlex
STD : 1200, 2400, 4800, 9600, 19200.
The baud rate is selected in the menu, Configuration / System / Inter-
face / Baud rate.

4.7.3 Selecting the output format of datasets

For downloading data to the interface you can select the output format.
It is selected in the menu, Configuration / System / Interface / Output
format.
The ASCII output format delivers formatted datasets.
The CSV output format delivers datasets separated by ";".
pHotoFlex Ser. no. 12345678
01.12.11 09:56:20
ID 1/ PROG 83
1.74 mg/l NH4-N [BV]
________________________________________
03.12.11 14:01:57
ID 2/ PROG 81
176 mg/l O2 [BV]
________________________________________
etc...