Total system monitoring (AQA2)

9.3.1

Entering standards

The standards compiled in the table "Spectroquant
CombiCheck and standard solutions" (see part 1
General information of this operating manual) are
already stored in the photometer. These values can
be overwritten.
For system monitoring, only one standard per test can
ever be stored.
The input of a standard is only complete when the
tolerances for finding it again have been input, i. e. it is
then first stored (no premature quitting).
9 Analytical Quality Assurance (AQA)
®
→ In the setup menu, call up the meter setup sub-
menu
→ Call up the AQA configuration submenu
→ Input the password
→ Call up the AQA standards submenu and the follo-
wing display appears:
→ Call up the standard solutions submenu
Select between
input
Input standards
output
Print/display standards
erase
Erase standards.
Displays the last selected method.
→ Select the method and confirm it
→ Input the standard.