9.1.3

Changing AQA intervals

The AQA intervals specify the time interval between
two AQA checks. Two separate intervals can be set up
for both photometer and system monitoring.
After an interval has expired, the following conse-
quences become effective:
Warning and loss of AQA identification
Locking of the method for concentration measure-
ments (as long as the locking is active).
Setting ranges:
Photometer monitoring (AQA1):
1 to 52 weeks (default: 12 w)
Total system monitoring (AQA2)
1 to 52 weeks (default: 4 w)
9 Analytical Quality Assurance (AQA)
→ In the AQA configuration menu, call up the AQA
intervals submenu.
→ Return without making any changes by pressing
→ Input the time interval for AQA meter via the
numeric keypad or using
→ Input the time interval for AQA system via the
numeric keypad or using
three times
, confirm using
, confirm using
.
49