10.1.3 Measuring the baseline

This function is used to newly determine the reference
absorbance or transmission for a wavelength range.
The baseline is stored and is used as a new reference
value for the spectrum measurement.
After the measurement is finished:
The measured baseline remains valid for all
the following spectrum measurements in the
same wavelength range and with the same
cell type.
Erasing the baseline
The baseline can be erased by
using the menu item, erase baseline
switching off the meter.
10 Special functions
→ Select the measure baseline function from the
spectrum menu.
→ Insert the cell to start the measurement.
The message measuring... appears on the display.
The baseline has been measured and stored.
→ Return to the spectrum menu using
→ Select the measure spectrum function from the
spectrum menu.
→ Measure as described in section
M
10.1.2
EASURING THE SPECTRUM
message measuring... is displayed, Baseline: OK
appears in the upper display line to indicate that
the measurement is being conducted against a
new baseline.
.
. While the
71