13 User-defined methods
Storing the characteristics:
The method data display is used to view
the data of the new user-defined method
again. The photometer stores the new data
only after
Sample display of the method data:
94
has been pressed.
27:05:99
user-def. methods
code no.
designation
wavelength
unit
citation form
resolution
zero point:
slope
MRB
MRE
reference cell
AQA theor. value
AQA tolerance
Starting point: table:
→ Storing the characteristics:
return: Pressing
and characteristics as a list.
→ Terminating without storing:
Use
or
to switch to the measuring
mode.
→ Scroll using
.
→ To store the new method and return, press
The method data is automatically printed.
11:37:43
306
TEST1
585
mg/l
COD
10
+0.090
+0.000496
100
1500
round
750
80
displays the code number
.