Appendix; Error Messages; Displayed Error Messages; Using The Xe Edp Command - Rice Lake IQ plus 350 Installation Manual

Digital weight indicator version 1.0. revolution scaleware configuration utility
Table of Contents

Advertisement

7.0

Appendix

7.1

Error Messages

The IQ plus 350 indicator provides a number of error messages. When an error occurs, the message is shown on
the indicator LCD display. Error conditions can also be checked remotely by using the XE EDP command as
described in Section 7.1.2.
Error Message
EA/D
A/D physical error
EEEROM
EEPROM physical error
EVIREE
Virgin EEPROM
EPCKSM
Parameter checksum error
EACKSM
A/D calibration checksum error
EFCKSM
Printer format checksum error
ELCKSM
Load cell calibration checksum error
EIDATA
Internal RAM checksum error
EREF
A/D reference error
OVERFL
Overflow error
------
Gross > overload limit
______
A/D underrange
7.1.1

Displayed Error Messages

The IQ plus 350 provides a number of front panel
error messages to assist in problem diagnosis.
Table 7-1 lists these messages and their meanings.
7.1.2

Using the XE EDP Command

The XE EDP command can be used to remotely query
the IQ plus 350 for the error conditions shown on the
front panel. The XE command returns two 5-digit
numbers in the format:
xxxxx yyyyy
where
contains a decimal representation of any
xxxxx
existing error conditions as described in Table 7-2.
If more than one error condition exists, the number
returned is the sum of the values representing the error
conditions. For example, if the XE command returns
the number 1040, this value represents the sum of an
A/D reference error (1024) and an A/D calibration
checksum error (16).
The second number returned (
assignments as shown in Table 7-2 to indicate whether
the test for the error condition was run. For example,
the value
= 50303 represents the decimal
yyyyy
equivalent of the binary value 1100 0100 0111 1111.
Using the bit assignments in Table 7-2, this indicates
all tests were run except the A/D physical error check.
34
IQ plus 350 Installation Manual
Description
Call Rice Lake Weighing Systems (RLWS) Service.
Use TEST menu to perform DEFLT (restore defaults) procedure, then
recalibrate load cells.
A/D converter requires recalibration. Call RLWS Service.
Call RLWS Service.
Recalibrate load cells.
Call RLWS Service.
A/D converter requires recalibration. Call RLWS Service.
Weight value too large to be displayed.
Gross value exceeds overload limit. Check configuration.
A/D reading < –4 mV. Check scale for binding or damage.
Table 7-1. IQ plus 350 Error Messages
) uses the same bit
yyyyy
Solution
Error
Code
Description
0
No error
1
EEPROM physical error
2
Virgin EEPROM
4
Parameter checksum error
8
Load cell calibration
checksum error
16
A/D calibration checksum
error
32
Print format checksum error
64
Internal RAM checksum error 0000 0000 0100 0000
128
not assigned
256
not assigned
512
A/D physical error
1024 A/D reference error
2048 not assigned
4096 not assigned
8192 not assigned
16384 A/D underrange
32768 Gross > overload limit
Table 7-2. Error Codes Returned on XE Command
Binary Value
0000 0000 0000 0000
0000 0000 0000 0001
0000 0000 0000 0010
0000 0000 0000 0100
0000 0000 0000 1000
0000 0000 0001 0000
0000 0000 0010 0000
0000 0000 1000 0000
0000 0001 0000 0000
0000 0010 0000 0000
0000 0100 0000 0000
0000 1000 0000 0000
0001 0000 0000 0000
0010 0000 0000 0000
0100 0000 0000 0000
1000 0000 0000 0000