Calibration; Front Panel Calibration - Rice Lake IQ plus 350 Installation Manual

Digital weight indicator version 1.0. revolution scaleware configuration utility
Table of Contents

Advertisement

4.0

Calibration

The IQ plus 350 can be calibrated using the front panel, EDP commands, or the Revolution
configuration utility. Each method consists of the following steps:
Zero calibration
Entering the test weight value
Span calibration
Optional rezero calibration for test weights using hooks or chains.
The following sections describe the calibration procedure for each of the calibration methods.
CONFIG
XXXXXXX
FORMAT
WZERO
*CAL*
Display and edit
Display and edit
test weight value
zero calibration
A/D count value
4.1

Front Panel Calibration

To calibrate the indicator using the front panel, do the
following:
1. Place the indicator in setup mode (display
reads
CONFIG
) and remove all weight from
the scale platform. If your test weights require
hooks or chains, place the hooks or chains on
the scale for zero calibration.
2. Press
until the display reads
Figure 4-1). Press
(
).
WZERO
3. With
displayed, press
WZERO
zero. The indicator displays
calibration is in progress. When complete, the
A / D c o u n t f o r t h e z e r o c a l i b r a t i o n i s
displayed. Press
calibration value and go to the next prompt
(
).
WVAL
4. With
displayed, place test weights on
WVAL
the scale and press
value. Use the procedure shown in Figure 4-2
to enter the actual test weight, then press
save the value and go to span calibration
(
WSPAN
).
5. With
WSPAN
displayed, press
span. The indicator displays
calibration is in progress. When complete, the
A / D c o u n t f o r t h e s p a n c a l i b r a t i o n i s
displayed. Press
calibration value and go to the next prompt
(
REZERO
).
CALIBR
SERIAL
WVAL
WSPAN
*CAL*
Display and edit
span calibration
A/D count value
Figure 4-1. Calibration (CALIBR) Menu
(see
CALIBR
to go to zero calibration
to calibrate
while
*CAL*
again to save the zero
to show the test weight
to calibrate
*CAL*
while
again to save the span
PROGRM
XXXXXXX
PFORMT
REZERO
CAL*
*
Press Enter to
remove offset from
zero and span
calibrations
6. The rezero function is used to remove a
calibration offset when hooks or chains are
used to hang the test weights.
If no other apparatus was used to hang the
test weights during calibration, remove
the test weights and press
the CALIBR menu.
If hooks or chains were used during
calibration, remove these and the test
weights from the scale. With all weight
removed, press
function adjusts the zero and span
calibration values. The indicator displays
* C A L *
calibrations are adjusted. When complete,
the adjusted A/D count for the zero
calibration is displayed. Press
the value, then press
CALIBR menu.
7. Press
until the display reads
press
to exit setup mode.
to
0
0 0 0 0 0
When editing numeric values, press
digit selected. Press
value of the selected digit.
Press
to save the value entered and return to the level above.
Figure 4-2. Editing Procedure for Numeric Values
Scaleware
DIG IN
XXXXXXX
ALGOUT
XXXXXXX
to return to
to rezero the scale. This
w h i l e t h e z e r o a n d s p a n
to return to the
CONFIG
or
to change the
or
to increment or decrement the
Calibration
VERS
to save
, then
25