Reporting Commands; The Resetconfiguration Command; Parameter Setting Commands - Rice Lake IQ plus 350 Installation Manual

Digital weight indicator version 1.0. revolution scaleware configuration utility
Table of Contents

Advertisement

5.1.2

Reporting Commands

Reporting commands (see Table 5-2) send specific
information to the EDP port. These commands can be
used in both setup mode and normal mode.
Command
DUMPALL
List all parameter values
VERSION
Write IQ plus 350 software version
P
Write current displayed weight with units
identifier. See Section 7.2 on page 35 for
more information.
ZZ
Write current weight and annunciator
status. See Section 7.2 on page 35 for
more information.
S
Write one frame of stream format
Table 5-2. EDP Reporting Commands
5.1.3

The RESETCONFIGURATION Command

The RESETCONFIGURATION command can be
used to restore all configuration parameters to their
default values. Before issuing this command, the
indicator must be placed in test mode (press and hold
setup switch for approximately three seconds to show
TEST menu).
Command
GRADS
Graduations
ZTRKBND
Zero track band
ZRANGE
Zero range
MOTBAND
Motion band
OVRLOAD
Overload
DIGFLTR1
Digital filtering
DIGFLTR2
DIGFLTR3
DFSENS
Digital filter cutout sensitivity
DFTHRH
Digital filter cutout threshold
TAREFN
Tare function
28
IQ plus 350 Installation Manual
Function
Description
1–60 000
OFF, 0.5D, 1D, 3D
1.9%, 100%
1D, 2D, 3D, 5D, 10D, 20D, OFF
FS+2%, FS+1D, FS+9D, FS
1, 2, 4, 8, 16, 32, 64
2OUT, 4OUT, 8OUT, 16OUT, 32OUT, 64OUT, 128OUT
NONE, 2DD, 5DD, 10DD, 20DD, 50DD, 100DD, 200DD, 250DD
BOTH, NOTARE, PBTARE, KEYED
Table 5-3. CONFIG EDP Commands
This command is equivalent to using the DEFLT
function on the TEST menu. See Section 7.8 on
page 41 for more information about test mode and
using the TEST menu.
NOTE:
settings are lost when the RESETCONFIGURATION
command is run.
5.1.4

Parameter Setting Commands

Parameter setting commands allow you to display or
change the current value for a particular configuration
parameter (Tables 5-3 through 5-10).
Current configuration parameter settings can be
displayed in either setup mode or normal mode using
the following syntax:
command<ENTER>
Most parameter values can be changed in setup mode
only. Use the following command syntax when
changing parameter values:
command=value<ENTER>
where
is either a number or a parameter value.
value
Use no spaces before or after the equal (=) sign. If you
type an incorrect command, the display reads
Changes to the parameters do not take effect until you
exit setup mode.
Values
All load cell calibration
.
??