9.0

Appendix

9.1

Error Messages

The 480 indicator provides a number of error messages. When an error occurs, the message prompts on the indicator LED
display.
9.1.1

Displayed Error Messages

The 480 provides a number of front panel error messages to assist in problem diagnosis.
their meanings.
Error Messages
E A/D
A/D physical error
EEEROM
EEPROM physical error
EVIREE
Virgin EEPROM
EPCKSM
Parameter checksum error
EACKSM
A/D calibration checksum error
EFCKSM
Printer format checksum error
ELCKSM
Load cell calibration checksum error Recalibrate load cells
EIDATA
Internal RAM checksum error
E REF
A/D reference error
ERROR
Internal program error
OVERFL
Overflow error
========
Gross > overload limit
- - - - - -
Gross < 20d behind zero
RNGERR
GRADS > 100,000
WVAL > 100,000
EEPERR
EEPROM error
HINOFF
High offset
LINOFF
Low offset
NOBATT
No battery
EUCKSM
Configuration checksum
OIMLER
OIML parameter error
EE-ACC
Accumulator error
Shorting the excitation voltage shuts the excitation voltage off. The only way to restore excitation voltage is to cycle
Note
power.
Description
Call Rice Lake Weighing Systems Service at 800-472-6703
Use TEST menu to perform DEFLT (restore defaults) procedure then recalibrate load cells
A/D converter requires recalibration. Call Rice Lake Weighing Systems Service
Call Rice Lake Weighing Systems Service at 800-472-6703
Call Rice Lake Weighing Systems Service at 800-472-6703
A/D converter requires recalibration; call Rice Lake Weighing Systems Service
Check configuration; call Rice Lake Weighing Systems Service if unable to clear error by
cycling power or if error recurs
Weight value is too large to be displayed
Gross value exceeds overload limit; check configuration or signal input level;
overload can be caused by input signal > 45 mV or common mode voltage > 950 mV
Gross value is more than 20 divisions behind zero (OIML)
Only displays in configuration mode
Call Rice Lake Weighing Systems for service at 800-472-6703
Zero load at startup is more than initial zero range (INIZR) setting of calibration zero–remove
the extra load
Zero load at startup is less than initial zero range (INIZR) setting of calibration zero–add the
missing load
The RTC lost time/date tracking at previous power off state due to low battery or no battery
condition; the printer, accumulator and audit functions will fail to get time and date
NOTE: Refers to internal coin battery only, not the rechargeable battery option.
The checksum value of configuration has changed from that stored in memory
Parameter set incorrectly for use in the OIML mode;
Error with the accumulator such as attempting to display an accumulated value greater than
six digits
Table 9-1. 480 Error Messages
Table 9-1
Solution
Example: Primary units set for lb or oz
© Rice Lake Weighing Systems ● All Rights Reserved
Appendix
lists these messages and
61