Transmit Power Display - Siemens Acuson S1000 Instructions For Use Manual

Diagnostic ultrasound system
Table of Contents

Advertisement

Adjusting the Transmit Power during Contrast Agent Imaging
During contrast agent imaging, you can also increase or decrease the transmit power. The
transmit power range is from 0.10% to 100% for contrast agent imaging.
Rotate the [Power] control located below the touch screen clockwise to increase transmit
power. Rotate the [Power] control counterclockwise to decrease transmit power.
For systems without a touch screen, rotate the [Power] LED panel control clockwise to
increase transmit power. Rotate the [Power] control counterclockwise to decrease transmit
power.

Transmit Power Display

The transmit power range is from 0% to 100%. Selecting 100%, in combination with other
ultrasound system controls or functions, generates the maximum acoustic intensity and
mechanical index for each transducer, where:
2
I
: ≤ 720 mW/cm
and MI ≤ 1.9
SPTA.3
I n s t r u c t i o n s f o r U s e
2 S a f e t y a n d C a r e
2 - 1 7