Adjusting the Transmit Power during Contrast Agent Imaging
During contrast agent imaging, you can also increase or decrease the transmit power. The
transmit power range is from 0.10% to 100% for contrast agent imaging.
●
Rotate the [Power] control located below the touch screen clockwise to increase transmit
power. Rotate the [Power] control counterclockwise to decrease transmit power.
○
For systems without a touch screen, rotate the [Power] LED panel control clockwise to
increase transmit power. Rotate the [Power] control counterclockwise to decrease transmit
power.
Transmit Power Display
The transmit power range is from 0% to 100%. Selecting 100%, in combination with other
ultrasound system controls or functions, generates the maximum acoustic intensity and
mechanical index for each transducer, where:
2
I
: ≤ 720 mW/cm
and MI ≤ 1.9
SPTA.3
I n s t r u c t i o n s f o r U s e
2 S a f e t y a n d C a r e
2 - 1 7