Siemens Acuson S1000 Instructions For Use Manual page 254

Diagnostic ultrasound system
Table of Contents

Advertisement

1 0 V i r t u a l T o u c h A p p l i c a t i o n s
To
Adjust the range of velocities
or elasticity displayed in the
region of interest
Activate the shadow function
in a dual-screen format
1 0 - 2 0
Do This
1. To change the maximum velocity or elasticity:
● Rotate [Max Velocity] or [Max Elasticity].
○ For systems without a touch screen, press and then rotate [Max Velocity].
2. To change the minimum velocity:
● Rotate [Min Velocity] or [Min Elasticity].
○ For systems without a touch screen, press and then rotate [Min Velocity] or
[Min Elasticity].
The system updates the shear wave velocities in the region of interest and the
color or gray bar to depict the velocities within the selected range. The selected
maximum and minimum velocity values display above and below the gray bar or
color bar. The selected maximum and minimum elasticity values display in blue.
1. When the measurement function is active, select Shadow.
2. Perform a measurement on either image.
When the system completes a shear velocity measurement, both images display
the measurement marker. When the system completes a 2D-mode measurement,
the measured image displays the measurement markers and the dotted line; the
shadow image displays only the dotted line.
3. To exit the shadow function, select Shadow.
I n s t r u c t i o n s f o r U s e