To remove and clean the back panel air filter:
Caution: Do not scrub, stretch, or bend the filter, or apply heat to the filter, as doing so could
damage the filter.
1.
Power off and unplug the power cord from the power outlet.
2.
Locate the air filter tray in the back panel and pull the tray out of the ultrasound system.
3.
Remove the air filter from the tray.
4.
Rinse the air filter with running water and allow the filter to completely dry.
To hasten drying, you may gently shake the filter, or blot the filter with a clean, lint-free
cloth.
Caution: Do not insert a wet filter as this can damage the system.
5.
Replace the air filter into the tray, positioning the filter flat side down, with the angled
corners fitting into the back of the tray. The filter will fit into the back corners of the tray only
if the filter is positioned flat side down.
6.
Press the filter against the hook and loop fasteners on the tray.
7.
Slide the air filter tray back into the ultrasound system.
8.
Plug the power cord into the power outlet.