Siemens Acuson S1000 Instructions For Use Manual page 62

Diagnostic ultrasound system
Table of Contents

Advertisement

2 S a f e t y a n d C a r e
To
Clean and disinfect the
trackball
2 - 3 2
Do This
Caution: Do not drop or place foreign objects inside the trackball assembly
because doing so may affect the trackball operation and damage the ultrasound
system.
Example of the trackball ring.
1
Tab for inserting into the trackball
assembly
2
Raised markings on either side of the
ring
1. Rotate the ring around the trackball counterclockwise and then carefully lift up the
ring to remove the ring and trackball.
2. Clean the ring and trackball with a cotton swab or lint-free pad moistened with mild
detergent solution.
3. Clean the inside of the trackball assembly with a cotton swab or lint-free pad
moistened with mild detergent solution.
4. As required, use an approved disinfectant wipe to disinfect the ring, trackball, and
trackball assembly.
5. Allow the trackball components to completely dry before reassembly.
6. Reinstall the ring and trackball.
a. Place the trackball inside the trackball assembly.
b. Place the ring over the trackball, taking care to align the tab with the point of
attachment on the trackball assembly.
c. Rotate the ring clockwise until the ring snaps into place and the ridges of the ring
are in a horizontal position.
Example trackball assembly.
1
Point of attachment for the trackball
ring
I n s t r u c t i o n s f o r U s e