Settings
Example
4C1
Abdomen
General
DVR Track 1 0:00:08
TIS: 0.4
TIB: 0.4
TIC: 1.0
MI: 1.2
MIF: 0.02
16fps
2D-mode Settings
Example
2D - 100%
THI
H5.50 MHz
I n s t r u c t i o n s f o r U s e
Parameter
Transducer name
Exam
 Image optimization preset for the active priority mode.
 During contrast agent imaging, indicates the balance of the image shown on the right
side of the screen:
– 2D displays only the 2D image
– CA displays only the image with the contrast agent
– MIX displays a combination of the 2D image and the image with contrast agent
(Displayed only when the DVR and media are ready for recording)
DVR track and time
Soft tissue thermal index
(Displayed only for fetal applications)
Bone thermal index
Cranial thermal index
Mechanical index
(Displayed only when contrast agent imaging is activated)
Maximum of the mechanical indices measured at the active focal zones
Frame rate
Parameter
Imaging mode / transmit power
Submode:
 THI indicates tissue harmonic imaging
 GEN indicates fundamental imaging
 Contrast indicates contrast agent imaging
 CCAI indicates contrast agent imaging (for systems without a touch screen)
Transmit frequency
Note: The transmit frequency is preceded by an H (harmonic) or HC (harmonic
compounding) when THI is activated. The transmit frequency is preceded by CPS or CHI
during contrast agent imaging when contrast pulse sequence or contrast harmonic
imaging is activated. ADI indicates the agent detection imaging technique is enabled.
4 E xa m i n a t i o n F u n d a m e n t a l s
4 - 2 1