Transducer Surface Temperature Limits

The following table provides the maximum surface temperature of the transducers compatible
with the system.
Maximum surface temperatures are in accordance with IEC 60601-2-37.
Transducer
4C1
6C1 HD
6C2
8C3 HD
9L4
12L4
14L5
14L5 SP
18L6 HD
4P1
4V1
4V1c
8V3
10V4
V5Ms
V7M
EV-8C4
EC9-4
MC9-4
7CF1
7CF2
9EVF4
CW2
CW5
AcuNav 8F
AcuNav 10F
TMM = Tissue Mimicking Material
NA = not applicable (not tested)
I n s t r u c t i o n s f o r U s e
Maximum Temperature
TMM
Still Air
≤ 38.6°C
≤ 35.9°C
≤ 41.0°C
≤ 38.0°C
≤ 40.1°C
≤ 40.0°C
≤ 41.0°C
≤ 42.5°C
≤ 39.4°C
≤ 38.1°C
≤ 42.5°C
≤ 47.0°C
≤ 41.7°C
≤ 41.1°C
≤ 40.7°C
≤ 40.5°C
≤ 42.9°C
≤ 43.0°C
≤ 40.9°C
≤ 42.6°C
≤ 40.3°C
≤ 40.8°C
≤ 42.3°C
≤ 46.0°C
≤ 41.4°C
≤ 41.0°C
≤ 37.7°C
≤ 36.5°C
≤ 36.4°C
≤ 39.4°C
≤ 42.6°C
≤ 40.3°C
≤ 38.9°C
≤ 39.6°C
≤ 36.0°C
≤ 36.7°C
≤ 42.5°C
≤ 45.3°C
≤ 38.6°C
≤ 45.0°C
≤ 36.7°C
≤ 37.2°C
≤ 38.4°C
≤ 38.3°C
≤ 36.9°C
≤ 37.7°C
≤ 35.3°C
≤ 37.7°C
NA
≤ 41.1°C
NA
≤ 39.6°C