During the Move

Caution: When moving the ultrasound system, protect it from environmental changes including:
moisture, winds, dirt and dust, and extreme heat or cold exposure.
Caution: Avoid moving the ultrasound system on outside surfaces with loose dirt, contaminates,
or standing liquids.
Caution: Care should be taken to minimize shock and vibration of the ultrasound system. Avoid
uneven surfaces that contain an abrupt height change or jarring surface irregularities.
You can move the ultrasound system from room to room within a facility and easily reposition
the system during an examination. Be careful on inclines and uneven surfaces. The ultrasound
system can be moved across pavement and other hardened parking lot surfaces.
Note: The wheels of the ultrasound system must be locked when transporting by vehicle. The
ultrasound system must be sufficiently anchored to the vehicle floor or walls such that it does not shift
or move during transport.

After the Move

Caution: Make sure the ultrasound system has proper ventilation during operation. Do not
position the system against walls or hard surfaces that would impede free ventilation around the
system.
Caution: Do not allow linens, bedding, and/or hanging curtain partitions to block the ultrasound
system's ventilation.
Caution: Obstructed fans can cause potential system overheating, system performance
degradation, or failure.
Caution: Brakes are most effective on a level surface. Never park the system on an incline
greater than five degrees.
Position system: Make sure the system is not placed against walls or fabrics that obstruct
perimeter air flow to the system cooling fans.
Set brakes: Set the front and rear brakes.
Unlock the flat panel monitor.
Plug in cord: Plug the power cord into a hospital-grade or local equivalent wall outlet.
If applicable, also plug the Ethernet cable into an appropriate connector.
Power ON: Power on ( ) the ultrasound system.
Check display: After the boot-up sequence is complete, verify that the image display is
stable, that you can select a transducer, and that selections made on the control panel
respond to your selection.