Temperature Sensor Failure (V5Ms)

WARNING: Disconnect the transducer if the system displays the message indicating a
temperature monitor failure. Remove the transducer from the patient. Failure to disconnect and
remove a transesophageal transducer with faulty temperature sensor can burn the patient's
esophagus. Contact your Siemens service representative.
Follow the instructions included in the on-screen message if the system detects a temperature
monitor failure:
Message
Temperature Monitor Failure
Temperature monitor failure.
Disconnect transducer.
I n s t r u c t i o n s f o r U s e
Displays When
The system detects a
temperature sensor
failure.
6 T r a n s e s o p h a g e a l T r a n s d u c e r
Recommended Action
1. Click Close to remove the message.
2. Disconnect the transducer.
3. Remove the transducer from the patient.
4. Contact your Siemens service
representative.
6 - 2 3