V7M Transducer Technical Description

Transducer Type:
Elements:
Range of Array Rotation
Angle:
Distal Tip Articulation:
Minimum Patient Weight:
Maximum insertion length:
Maximum insertion width:
2D-mode Transmit
Frequencies:
Color mode Transmit
Frequencies, including
CDV, Color DTI:
Doppler Transmit
Frequencies, including PW,
SCW, Aux CW,
Pulsed Wave DTI:
Imaging Modes:
Maximum Image
Field of View:
Image:
Operating Environment
Temperature:
Relative Humidity:
Storage Environment
Temperature:
Relative Humidity:
Degree of protection
against electrical shock:
Degree of protection
against ingress of fluid:
I n s t r u c t i o n s f o r U s e
Phased Array, multiplane
48
-12° to 192°
120° anterior, 90° posterior
3.5 kg (7.7 lbs) and above
700 mm
10.7 mm
4.0 MHz, 5.0 MHz, 6.0 MHz, 7.0 MHz, 8.0 MHz
3.5 MHz, 3.75 MHz, 4.0 MHz, 4.75 MHz
3.5 MHz, 3.75 MHz, 4.0 MHz
2D-mode, PW Doppler, Steerable CW Doppler, Color mode, Color DTI,
Pulsed Wave DTI, M-mode, Anatomical M-mode
90°
Sector
0°C to +45°C
10% to 95%
-10°C to +45°C
10% to 95%
V7M transducers with this label are certified as type BF (not
defibrillator proof) per IEC 60601-1 Standard for Safety of
Medical Equipment
IPX8
6 T r a n s e s o p h a g e a l T r a n s d u c e r
6 - 3 9