6 T r a n s e s o p h a g e a l T r a n s d u c e r
Measurement Accuracy
Rotation angle of the
transducer array:
Temperature:
EMC Note: Operating the transducer in close proximity to sources of electromagnetic fields, such as
radio transmitters, may lead to temporary degradation or interference visible on the image monitor
screen. A lightening of image background may be noticed while visualizing hypoechoic structures, or
color spectral interference, or jitter, or horizontal lines in the image screen may occur. The transducer
and system have been designed and tested to withstand such interference and will not be
permanently affected.
6 - 3 8
≤3°
Temperature
>41°C
>=41°C
>=42°C
>=42°C and system freeze
Accuracy
40.1 ≤ 41°C
40.1 ≤ 41°C
41.1 ≤ 42°C
41.9 ≤ 43°C
I n s t r u c t i o n s f o r U s e