Autotolerances Setting; Figure 56. Hitch Standardization - HunterLab EasyMatch User Manual

Hide thumbs Also See for EasyMatch:
Table of Contents

Advertisement

Table 4. Hitch Standardization By Additive or Ratio Application
Hitch Standard
Target
Measured
X
80.27
78.29
Y
81.00
79.21
Z
50.71
47.76

AutoTolerances Setting

Using Tab 4, AutoTolerances are calculated for a Color Scale using CMC. The default
values of l:C – 2:1 with auto correction factor = 0.75 and commercial factor = 1.
However, these ratios can be modified as needed.
User's Manual for Aeros and EasyMatch Essentials v 1.1
3. The Standard is displayed under As Read. Enter the target values.
4. Check the Apply Hitch box and press OK.

Figure 56. Hitch Standardization

Calculate
Additive
Hitch
= +1.98
= + 1.79
= +2.95
Hitch Factor
Hitch Factor
1.98
1.025290586
1.79
1.022598156
2.95
1.061767169
Calculate
Ratio
Hitch
New Read
=*1.025
70.84
=*1.022
72.25
=*1.061
46.07
Apply Hitch
Apply Hitch
Additive
Ratio
New Read
New Read
72.82
72.63
74.04
73.88
49.02
48.91
Hitch Calc
Hitch Calc
70.84+1.98
=72.84*1.025
72.25+1.79
=72.25*1.022
46.07+2.95
=46.07*1.061