Figure 44. Slope & Bias Correction; Figure 45. Input Gain & Bias - HunterLab EasyMatch User Manual

Hide thumbs Also See for EasyMatch:
Table of Contents

Advertisement

User's Manual for Aeros and EasyMatch Essentials v 1.1
Gain Correction= (Expected Value 1-Expected Value 2)/
(Measured Value 1- Measured Value 2)
3. Linear regression: Create a y=mx + b relationship comparing actual
readings to target values, where target values is on the Y-axis and actual
readings are on the x-axis. Enter the slope correction under Gain and the
intercept correction under Bias.
Select Index
Bias
Configuration
Figure 44. Slope & Bias Correction
Figure 45. Input Gain & Bias
The indices with bias correction include: 457nm Brightness, BCU, HCCI,
SCAA, Tint E313, WI E313, Y Brightness, YI D195, YI E313, Z%.
36