Operating Instructions
Adapters for Specimen Clamping
Changing and/or Fastening
Specimen Clamps
44
Operation Manual HM335S
The available specimen clamps are all fastened or removed in the same
way.
[b]
•
To change the specimen clamping system, press the clamping lever
[a] downwards and pull it to the side.
•
Now the specimen clamp can be pulled to the front and another
specimen clamp can be placed into the clamping system.
•
Insert the new specimen clamp into the cylinder head so that the
clamping lever [a] can be put through the hole of the orienting
adapter [d] and pulled through from the right side.
•
Then align the specimen clamp with the orienting
screws [b] in the X- and Y-axis and orientate it on the Z-axis.
•
Press the clamping lever upwards.