Changing And/Or Fastening Specimen Clamps - Thermo Scientific Microm HM355S Operation Manual

Rotary microtome
Hide thumbs Also See for Microm HM355S:
Table of Contents

Advertisement

Operating Instructions
Adapters for Specimen Clamping
Changing and/or Fastening
Specimen Clamps
44
Operation Manual HM335S
The available specimen clamps are all fastened or removed in the same
way.
[b]
To change the specimen clamping system, press the clamping lever
[a] downwards and pull it to the side.
Now the specimen clamp can be pulled to the front and another
specimen clamp can be placed into the clamping system.
Insert the new specimen clamp into the cylinder head so that the
clamping lever [a] can be put through the hole of the orienting
adapter [d] and pulled through from the right side.
Then align the specimen clamp with the orienting
screws [b] in the X- and Y-axis and orientate it on the Z-axis.
Press the clamping lever upwards.
[a]
[d]