Large Display Mode
Turning off the
"retraction" Function
Thermo Fisher Scientific
If desired, the function "Retraction" can be turned off.
Press the MENU button [1]
Select submenu "Retraction" by turning the knob [A].
Press the knob to confirm the selection.
Turn the knob [A] to select the desired function: ON or OFF
Press the knob [A] to confirm the setting.
The yellow LED RETRACTION might stay on and will go off only
after the specimen has been passed through the cutting movement by
turning hand wheel.
To turn the „Retraction" mode on again, please proceed as descri-
bed above.
Note!
The selected function is shown on the display as ON or OFF.
Operating Instructions
Display and Key Functions
Operation Manual HM335S
37