Continuous Stroke
Thermo Fisher Scientific
Multi Stroke
Press the MENU button [1].
Select the function "Multi" in the submenu "Mode" by turning the
knob [C] and confirm by pressing it.
Then press the knob [C] again.
Then turn it to select the desired number of sections. The number
of the cutting strokes depends on the chosen pre-selection
(2 up to 99).
Unlock the lever for mechanical stroke.
Press twice the START/STOP button [7] or twice the knob [C] or
twice the foot pedal (optional), to release a multiple cutting cycle.
After the set number of sections has been carried out, the movement
stops in the upper reversal point.
Press the MENU button [1].
In the submenu "Mode" select the function "Cont" via the knob
[A]or button [8].
Unlock the lever for the mechanical brake.
To start a continuous cutting cycle, press twice the START/STOP
button or twice knob [C] or step twice on the foot pedal (optional).
To stop the continuous stroke, press knob [C] or the brake button [4]
or the foot pedal (optional) once more.
Note!
If a cutting cycle has been started, it will run through to the end and stop
in the next upper reversal point.