Standard Specimen
Thermo Fisher Scientific
The unrestricted use of the Macro-UCC is only possible with ER blade
carrier.
Note!
Maximum 68mm x 50mm specimen size can be cut. The Paraffin blocks in
MacrOflow-Cassettes are 70-72mm long and should be trimmed to 68 mm
length manually.
The standard specimen clamp is used for rectangular and square paraffin
and plastic blocks.
Clamp
Insert the specimen against the fixed jaw [b] first.
Then tighten with the clamping screw [a] to tighten the specimen
via the movable jaw [c].
Note!
For the stability of the specimen, do not let it project too much over the
clamping jaws.
Note!
The standard specimen clamp is equipped with two different movable jaws
[c], which are of different weight. The lightweight jaw is used with inserts
for round specimens.
[a]
[a]
Operating Instructions
Specimen Clamping
[b]
[c]
[a]
[b]
[c]
Operation Manual HM335S