Section 2
Operation

Platform Stalled

Check Fuse

Turning Audible
Alarm Off and On
2-8
Orbital Shakers
Platform Stalled alerts the operator that
something is preventing free platform
movement. The motor automatically shuts
off and the audible alarm, screen message and warning lights are initiated.
The motor will attempt restart after approximately 15-20 seconds.The
motor will continue to cycle On and Off until the obstruction is removed
or the unit is turned off. On motor restart, the audible alarm and warning
lights are automatically cleared. The screen message remains until cleared
by the operator.
Check Fuse alerts the operator that the
primary drive motor fuse has blown. The
audible alarm, screen message and
warning lights are initiated. When the unit is turned on after fuse
replacement, all alarm indicators are cleared.
To access the system Configuration menu, press the down arrow, the up
arrow and the Silence button in that sequence.This screen will appear on
the display.