Aethalometer
Model AE33
10.1.4 TSI_4100
The TSI-4100 is a mass flowmeter. It may be used for Automatic calibration, in which case it
must be connected to the COM3 port. Its flow data is stored in the 'Fields_i' section.
10.1.5 Datalogger_AE33_protocol
Please check the table in section 10 for more details.

10.1.6 Sample Stream Dryer

The Sample Stream Dryer transmits the following data: temperature, RH and dew point at the
inlet; and temperature, RH and dew point at the outlet. This data is stored in the 'Fields_i'
section.

10.1.7 "Bayern Hessen" datalogger protocol

The instrument supports data acquisition using the BH protocol. The basic command is:
<STX>DA<address><ETX><BCC>. This command initiates a data transfer from the instrument.
The address (a value between 1 and 255) is optional and can be set on the 'Operation /
Advanced' screen as "External ID".
The instrument responds to a valid command in the following format:
<STX>MD07<SP><address><BC1><SP><opStatus><SP><errStatus><SP>0000000000<SP>....
....
<address><BC7><SP><opStatus><SP><errStatus><SP>0000000000<SP><ETX><BCC><CR><LF>
<address> is the instrument address.
<BC1> ... <BC7> are 7 BC values, shown in exponential representation with 4 characters
mantissa and 2 characters exponent.
<opStatus> is the status of the operation state:
00000000 –measuring,
00000001 – tape advance,
00000010 – ATN_zero measurement,
00000011 – instrument stopped,
____________________________________________________________________________________________________________
User's manual – Ver. 1.55
July 2017
95/146