Aethalometer
Model AE33
4. A new screen will appear. Choose "Manual". The "Auto" option can be only used if you are
using the TSI 4110 mass flow meter connected by data cable to the AE33.
5. The instrument will first measure the flow through the filter tape.
6. Confirm by pressing "OK" button.
7. Connect your external mass flow meter to the inlet port.
8. Connect the flow meter using well-sealed tubing and fittings. Make sure that the
connections are tight.
10. Press "OK".
11. Choose the flow at which you want to perform the test. Normally this test is conducted at 5
LPM. Press "OK"
____________________________________________________________________________________________________________
User's manual – Ver. 1.55
July 2017
72/146