14. Press the input box field and enter the flow from the external flow meter.
IMPORTANT ! – units are milliliters per minute (mLPM).
15. Confirm by pressing "Enter".
16. After entering the last value (@ 5LPM) and pressing "OK", the procedure is completed.
17. When finished, remove the flow calibration pad and replace the filter tape. Press "OK"
when done.
18. The flow calibration procedure is now completed.